The use of EBT3 film and Delta4 for the dosimetric verification of EclipseTM treatment planning system in a heterogeneous chest phantom: An IMRT technique

Abedi Firouzjah, R. and Nickfarjam, A. and Bakhshandeh, M. and Farhood, B. (2019) The use of EBT3 film and Delta4 for the dosimetric verification of EclipseTM treatment planning system in a heterogeneous chest phantom: An IMRT technique. International Journal of Radiation Research, 17 (2). pp. 355-361.

[img] Text
IJRR.pdf

Download (1MB)
Official URL: https://www.scopus.com/inward/record.uri?eid=2-s2....
DOI: UNSPECIFIED

Abstract

Background: This study aimed to evaluate the dose calculation accuracy of EclipseTM treatment planning system (TPS) in a heterogeneous chest phantom with the intensity modulated radiotherapy (IMRT) technique using EBT3 film and Delta4. Materials and Methods: Two IMRT plans (A and B) were prepared for radiotherapy of the heterogeneous chest phantom. Plan A was between the left lung and the surrounding soft tissue and plan B was carried out on the soft tissue. EBT3 film and Delta4 were used for dose measurement in the phantom. EclipseTM TPS was also used for dose calculation. Finally, the gamma index values of the TPS with film and TPS with Delta4 were obtained. A 95 passing rate of gamma index with the passing criterion of 3mm/3 and a dose threshold of 20 as the standard criterion was considered in this study. Furthermore, the passing rates of gamma indices of the film and Delta4 were compared with each other via Bland-Altman analysis. Results: The mean passing rate of gamma index with standard passing criterion between the TPS calculations and film measurements was 96.95±0.22, while it was equal to 97.7±0.56 and 98.45±0.21 between the TPS calculations and 2D and 3D Delta4 measurements, respectively. Additionally, the differences between the passing rates of gamma indices of the film and Delta4 were less than 5. Conclusion: The findings demonstrate that the accuracy of dose calculations of EclipseTM TPS in a heterogeneous chest phantom with the IMRT technique is within the standard passing criterion. Furthermore, it can be concluded that there is a good agreement between the film and Delta4, as IMRT QA devices. © 2019 Novin Medical Radiation Institute. All rights reserved.

Item Type: Article
Additional Information: cited By 0
Uncontrolled Keywords: anisotropic analytical algorithm; Article; calibration; controlled study; dosimetry; intensity modulated radiation therapy; intermethod comparison; left lung; measurement accuracy; radiotherapy dosage; treatment planning; x-ray computed tomography
Subjects: Medicine
Divisions: Faculty of Para medicine > Department of Management Radiology and Medical Physics
Depositing User: ART . editor
Date Deposited: 29 Dec 2019 12:51
Last Modified: 29 Dec 2019 12:51
URI: http://eprints.kaums.ac.ir/id/eprint/4730